Multiplexer OPMUX-01/4

Technical data
Channels: 4*
Pulse amplitude range: 50-360V*
Max cable length between
computer and the box: 30m
Connectors BNC:
Transducers Trigger input
Signal output
DB15:
Control signals/Power supply
* - software selected
Features and Specifications
Pulser circuit waveforms: The rising edge of the Trig In signal initiates the transducer charging process which takes about 3us. After this time a transistor switch which discharges the transducer is turned on (discharge time is about 20ns, but it can be longer, if the capacitance of the transducer is to large; the limit is reached, if the transducer is made from the standard ceramics, is 0.1 mm thick and has 8 mm diameter).
Because of the very low output impedance of the device (<1 Ω) and short discharge time the pulse generated with this
device could be concerned as a real pulse answer for the most transducers.
Need a solution for a specific measurement problem?
OPTEL can evaluate the measurement physics, transducers, electronics and software before a full system is built.
Start with a feasibility studyContact OPTEL